UNKNOWN_RECONSTRUCTIVE_PRODUCT
Report
- Report Number
- 0002249697-2014-03750
- Event Type
- Injury
- Date Received
- October 3, 2014
- Date of Event
- August 20, 2014
- Report Date
- September 7, 2014
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- KWA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
CATALOG NUMBER IS UNKNOWN AT THIS TIME. THE DEVICE WAS REPORTED AS AN UNKNOWN TRIDENT CUP. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT. NOT RETURNED.
THIS EVENT IS RELATED TO VOLUNTARY RECALL RA 2012-067. THIS RECALL WAS INITIATED DUE TO THE POTENTIAL RISKS ASSOCIATED WITH MODULAR NECK STEMS. THE DEVICE REPORTED IN THIS INVESTIGATION DID NOT CONTRIBUTE TO THE REPORTED EVENT.
DR (B)(6) AT THE CLINIC, REPORTED THE FOLLOWING EVENT: "PLEASE SAVE THIS NEW VIGILANCE CASE MATERIAL ON STRYKER ABGII MODULAR (B)(4). PATIENT FIRST IMPLANTED ON (B)(6) 2011: TOTAL HIP REPLACEMENT (RIGHT) - DUE TO HIP OSTEOARTHRITIS - WITH ABGII MODULAR STEM AND TRIDENT CUP. CONVOCATION OF THE PATIENT FOR CONTROL (AS RECOMMENDED BY THE ANSM MONITORING PROTOCOL). PAINFUL FOR A YEAR. OSTEOLYSIS WITH BONE LOSS OF MORE THAN 1 CM FROM MERKEL. TISSUE REACTION WITH EFFUSION ON ULTRASOUND. COBALT LEVEL 4.8 MICROG/L. REVISION SURGERY PERFORMED THE (B)(6) 2014 - BIPOLAR CHANGE AND MERKEL RECONSTRUCTION WITH BONE GRAFT. TYPICAL GREENISH APPEARANCE OF INTRA-ARTICULAR EFFUSION AND CAPSULAR INFILTRATION. MODULAR NECK CORROSION. INTRA ARTICULAR COBALT LEVEL = 4789 MICROG/L. UNEVENTFUL FOR NOW."
DR (B)(6), SURGEON AT THE CLINIC, REPORTED THE FOLLOWING EVENT : "PLEASE SAVE THIS NEW VIGILANCE CASE MATERIAL ON STRYKER ABGII MODULAR (SURGEON FILE REFERENCE: (B)(4)). PATIENT FIRST IMPLANTED ON (B)(6) 2011 : TOTAL HIP REPLACEMENT (RIGHT) - DUE TO HIP OSTEOARTHRITIS - WITH ABGII MODULAR STEM AND TRIDENT CUP. CONVOCATION OF THE PATIENT FOR CONTROL (AS RECOMMENDED BY THE (B)(4) MONITORING PROTOCOL). PAINFUL FOR A YEAR. OSTEOLYSIS WITH BONE LOSS OF MORE THAN 1 CM FROM MERKEL. TISSUE REACTION WITH EFFUSION ON ULTRASOUND. COBALT LEVEL 4.8 MICROG/L. REVISION SURGERY PERFORMED THE (B)(6) 2014 - BIPOLAR CHANGE AND MERKEL RECONSTRUCTION WITH BONE GRAFT. TYPICAL GREENISH APPEARANCE OF INTRA-ARTICULAR EFFUSION AND CAPSULAR INFILTRATION. MODULAR NECK CORROSION. INTRA ARTICULAR COBALT LEVEL = 4789 MICROG/L. UNEVENTFUL FOR NOW."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 618425 | UNKNOWN_RECONSTRUCTIVE_PRODUCT | IMPLANT | KWA | STRYKER ORTHOPAEDICS-MAHWAH | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Required Intervention |