FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 4140241 · Received October 3, 2014

Report

Report Number
1416980-2014-34384
Event Type
Malfunction
Date Received
October 3, 2014
Date of Event
September 8, 2014
Report Date
September 8, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. HOWEVER, AN INCOMPLETE PRIMED PATIENT LINE IS A KNOWN CAUSE OF THIS ALARM. ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE,¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE, PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET AND SAYS TO VERIFY THAT THE PATIENT LINE IS PROPERLY PRIMED. IT WARNS THE USER NOT TO CONNECT TO THE PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HOMECHOICE DEVICE EXPERIENCED A SYSTEM ERROR 2240 (AIR IN LINE/SET). THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE ALARM. THIS OCCURRED DURING THE INITIAL DRAIN OF PERITONEAL DIALYSIS (PD) THERAPY. THE PATIENT DID NOT CHECK TO VERIFY THAT THE PATIENT LINE WAS FULLY PRIMED BEFORE THEY CONNECTED. THE PATIENT WAS TO COMPLETE THERAPY BY STARTING OVER WITH NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
620676 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 45 YR HOMECHOICE