FDA Adverse Event Injury Summary report: N

ALTRUA

MDR report key: 4140092 · Received October 3, 2014

Report

Report Number
2124215-2014-17399
Event Type
Injury
Date Received
October 3, 2014
Date of Event
August 27, 2014
Report Date
October 27, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
PMA / PMN Number
P970003
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PRODUCT HAS NOT BEEN RETURNED FOR TESTING. THIS PRODUCT ISSUE WILL BE UPDATED IF ADDITIONAL DETAILS ARE RECEVIED.

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED FOR TESTING. THIS PRODUCT ISSUE WILL BE UPDATED WHEN EVALUATION IS COMPLETE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER AND RIGHT VENTRICULAR (RV) LEAD EXHIBITED NOISE, OVERSENSING AND PACING INHIBITION. A LEAD FRACTURE WAS SUSPECTED, BUT COULD NOT BE CONFIRMED. A REVISION PROCEDURE WAS PERFORMED AND THE SYSTEM WAS REMOVED FROM SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
620852 ALTRUA IMPLANTABLE PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND S603

Patients

Seq Age Sex Outcome Treatment
1 81 YR Hospitalization| L| R 4087| S603| 1296| 4088