FDA Adverse Event Malfunction Summary report: N

COGNIS

MDR report key: 4140011 · Received October 3, 2014

Report

Report Number
2124215-2014-17241
Event Type
Malfunction
Date Received
October 3, 2014
Date of Event
September 8, 2014
Report Date
October 13, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Removal / Correction Number
Z-0087-2015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON COMPLETION FROM ANALYSIS, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY. REVIEW OF THE DEVICE MEMORY INDICATED THAT A LOW VOLTAGE ALERT, CODE 1003, WAS RECORDED. THE BATTERY VOLTAGE WAS LOWER THAN EXPECTED, BUT STILL SUPPORTED FULL DEVICE FUNCTION. A SERIES OF AUTOMATED ELECTRICAL/FUNCTIONAL TESTS WERE CONDUCTED AND NO ISSUES WITH THE DEVICE'S ABILITY TO DELIVER THERAPY WERE OBSERVED; BASIC SENSING, PACING AND SHOCKING FUNCTIONS OF THE DEVICE WERE VERIFIED. USING HISTORICAL DAILY BATTERY VOLTAGE MEASUREMENT DATA, ENGINEERS DETERMINED THAT THIS DEVICE WAS DEMONSTRATING BEHAVIOR CONSISTENT WITH A HIGH CURRENT CONDITION ASSOCIATED WITH A COMPROMISED LOW VOLTAGE CAPACITOR CONNECTED TO THE DEVICE'S BATTERY. LOW VOLTAGE CAPACITORS ARE USED IN THE DEVICE'S HIGH VOLTAGE CHARGING OPERATION IN ORDER TO FACILITATE FAST CHARGE TIMES. MALFUNCTION OF THESE CAPACITORS RESULTED IN A HIGH CURRENT DRAIN, WHICH WAS DEPLETING THIS DEVICE'S BATTERY FASTER THAN NORMAL. BOSTON SCIENTIFIC HAS ISSUED AN ADVISORY COMMUNICATION REGARDING AN OLDER SUBSET OF COGNIS/TELIGEN DEVICES THAT IS MORE SUSCEPTIBLE TO THIS ANOMALY. SPECIFICALLY, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING AN INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS PARTICULAR DEVICE WAS INCLUDED IN THE ADVISORY POPULATION.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE REMAINS IN SERVICE AT THIS TIME. THIS REPORT WILL UPDATED UPON RECEIPT OF ADDITIONAL INFORMATION.

Description of Event or Problem · 1

THE DEVICE WAS SUBSEQUENTLY EXPLANTED, REPLACED AND RETURNED FOR LABORATORY ANALYSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

AS PART OF AN ANALYSIS TO REVIEW THE SAFETY ARCHITECTURE LOW VOLTAGE ALERT (CODE 1003), BATTERY VOLTAGE DATA WAS PULLED FROM THE REMOTE PATIENT MONITORING SYSTEM. DURING OUR ANALYSIS, THE DEVICE WAS IDENTIFIED AS DEMONSTRATING EVIDENCE OF PREMATURE BATTERY DEPLETION THAT COULD POTENTIALLY LEAD TO COMPROMISED THERAPY IF NOT ADDRESSED IN A TIMELY MANNER. BOSTON SCIENTIFIC TECHNICAL SERVICES PROACTIVELY REACHED OUT TO THE FIELD REPRESENTATIVE TO NOTIFY THEM OF THIS FINDING AND RECOMMEND DEVICE REPLACEMENT. AT THIS TIME, THE DEVICE REMAINS IMPLANTED AND NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
620246 COGNIS IMPLANTABLE CHF PULSE GENERATOR LWS GUIDANT CRM CLONMEL IRELAND N119

Patients

Seq Age Sex Outcome Treatment
1 72 YR N119| 4469| 4470| 1290| 4456| 4543| 0185