FDA Adverse Event Injury Summary report: N

ASANTE SNAP INSULIN PUMP SYS

MDR report key: 4135675 · Received September 26, 2014

Report

Report Number
3009351200-2014-00010
Event Type
Injury
Date Received
September 26, 2014
Date of Event
August 29, 2014
Report Date
September 26, 2014
Manufacturer
ASANTE SOLUTIONS INC.
Product Code
LZG
PMA / PMN Number
K122483
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

PT CODE: VOMITING. DEVICE CODE: FAILURE TO INFUSE, USE OF DEVICE ISSUE. METHOD CODE: LABELING EVALUATION (REVIEWING ALL LABELING (INCLUDING INSTRUCTIONS FOR USE, MANUALS, AND PACKAGING) FOR LANGUAGE RELATED TO REPORTED DEVICE PROBLEMS.)) RESULTS CODE: DEFORMATION PROBLEM. CONCLUSION CODE: OPERATIONAL CONTEXT CAUSED OR CONTRIBUTED TO EVENT. DEVICE NOT RETURNED. KNOWN INHERENT RISK OF PROCEDURE.

Description of Event or Problem · 1

ASANTE TERRITORY SALES MGR INFORMED CUSTOMER SERVICE THAT A (B)(6) YEAR OLD PT WAS IN THE NURSES OFFICE AT SCHOOL WITH A BLOOD GLUCOSE LEVEL OF 600 MG/DL AND WAS VOMITING. THE ISSUE APPEARED TO BE WITH THE CANNULA. THE PT HAD PREVIOUSLY EXPERIENCED THIS SAME ISSUE WHILE USING THE ANIMAS PUMP. ACCORDING TO THE ASANTE CLINICAL MGR WHO TRAINED THIS PT, THE ISSUE STEMS FROM PT BEING VERY LEAN AND ACTIVE (IS A CHEERLEADER) AND THE CANNULA IS KINKING SO SHE IS NOT GETTING THE INSULIN SHE NEEDS. LOCATION OF CANNULA WAS BUTTOCKS. NO ALARM WAS ACTIVATED WITH THE SNAP PUMP SYS. ACCORDING TO THE PT'S MOTHER, HER BLOOD GLUCOSE THE PREVIOUS NIGHT WAS 106 MG/DL, THE MORNING OF THE EVENT BEFORE BREAKFAST IT WAS 295 MG/DL. SHE BOLUSED AND WENT TO SCHOOL WHERE IT WAS THEN AT 500 MG/DL. ASANTE SENT PT EXTRA INFUSION SETS SINCE SHE IS GOING THROUGH THEM SO FAST. ASANTE DISTRIBUTES THE INFUSION SETS WHICH ARE MANUFACTURED BY ANOTHER COMPANY. THE MFR WAS INFORMED OF THIS COMPLAINT ON (B)(6) 2014. ASANTE IS REPORTING THIS EVENT BECAUSE OF OFF-LABEL USE OF THE SNAP INSULIN PUMP SYS IN A PEDIATRIC CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600217 ASANTE SNAP INSULIN PUMP SYS PUMP, INFUSION, INSULIN, 880.5725 LZG ASANTE SOLUTIONS INC. 7115

Patients

Seq Age Sex Outcome Treatment
1 16 YR Other| R