FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 4134014 · Received October 2, 2014

Report

Report Number
2939301-2014-26137
Event Type
Malfunction
Date Received
October 2, 2014
Report Date
September 27, 2014
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN USA ALLEGING THAT THE SUBJECT METER (ONETOUCH VERIO IQ) READ INACCURATE COMPARED TO ANOTHER METER. THE REPORTER WAS UNABLE TO GIVE EXACT VALUES. THE TESTS WERE PERFORMED WITHIN AN UNKNOWN TIME OF EACH OTHER. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING, SINCE WE COULD NOT CONFIRM THAT THE CALCULATED DIFFERENCE EXCEEDED OUR CRITERIA AS NO VALUES WERE GIVEN. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
616085 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3649814

Patients

Seq Age Sex Outcome Treatment
1