ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-28065
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Report Date
- September 19, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/19/2014 WITH THE FOLLOWING FINDINGS: THE PUMP POWERED ON WITH RETURNED BATTERY CAP. THE BATTERY CAP WAS ABLE TO FULLY TIGHTEN AND WAS MEASURED TO BE WITHIN SPECIFICATIONS. THERE WAS EVIDENCE OF MOISTURE CONTAMINATION INSIDE THE BATTERY COMPARTMENT AND ON THE UNDERSIDE OF THE BATTERY CAP. THERE WERE PUMP REBOOT EVENTS OBSERVED IN THE PUMP'S BLACK BOX ON (B)(6) 2014. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO REBOOT, LOSS OF POWER OR CALL SERVICE ALARMS DUPLICATED. NO "INTERMITTENT POWER" EVENTS WERE DUPLICATED DURING INVESTIGATION. THE PUMP PASSED A LEAK TEST AND THERE WAS NO ADDITIONAL EVIDENCE OF MOISTURE INSIDE THE PUMP.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (MOISTURE INGRESS) ISSUE. IT WAS REPORTED THAT EVIDENCE OF MOISTURE INGRESS WAS FOUND IN THE BATTERY COMPARTMENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 611619 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |