6.5CM ADULT CRANI ATTACHMT
Report
- Report Number
- 1045834-2014-13121
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Report Date
- April 3, 2013
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- PK133604
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL NARRATIVE: SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY TECHNICIAN, IT WAS NOTED THAT THE NEURO TIP WAS DAMAGED AS WELL AS THE CRANI LUG AND THE BALL BEARING WAS WORN DUE TO INAPPROPRIATE HANDLING. DEVICE WAS INSPECTED, AND REPAIRED BY REPLACING A FEW COMPONENTS.
CUSTOMER REPORTED THAT THE CONSOLE DISPLAYED THE HANDPIECE OVERHEAT WARNING, IMPENDING HANDPIECE SHUTDOWN ERROR CODE. THIS IS 1 OF 1 REPORT FOR EVENT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 612769 | 6.5CM ADULT CRANI ATTACHMT | MOTOR, DRILL, ELECTRIC | HBC | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |