CAD II
Report
- Report Number
- 8030965-2014-01464
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Report Date
- April 26, 2013
- Manufacturer
- SYNTHES GMBH
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE ENGINE WAS STARTING BY ITSELF DUE TO A CASE DEFORMED BY IMPACT. THIS WAS ATTRIBUTED TO NORMAL WEAR. THE DEVICE WAS RETURNED TO THE CUSTOMER ON (B)(6) 2013. PLACEHOLDER.
CUSTOMER REPORTED THE DRILL IS CONSTANTLY SPINNING, NO TRIGGER ACTIVATED. DEVICE WAS RETURNED FOR SERVICE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 612767 | CAD II | HWE | SYNTHES GMBH | 32133 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |