FDA Adverse Event Malfunction Summary report: N

CAD II

MDR report key: 4133431 · Received October 1, 2014

Report

Report Number
8030965-2014-01464
Event Type
Malfunction
Date Received
October 1, 2014
Report Date
April 26, 2013
Manufacturer
SYNTHES GMBH
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE ENGINE WAS STARTING BY ITSELF DUE TO A CASE DEFORMED BY IMPACT. THIS WAS ATTRIBUTED TO NORMAL WEAR. THE DEVICE WAS RETURNED TO THE CUSTOMER ON (B)(6) 2013. PLACEHOLDER.

Description of Event or Problem · 1

CUSTOMER REPORTED THE DRILL IS CONSTANTLY SPINNING, NO TRIGGER ACTIVATED. DEVICE WAS RETURNED FOR SERVICE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).