FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4133373 · Received October 1, 2014

Report

Report Number
2032227-2014-31385
Event Type
Malfunction
Date Received
October 1, 2014
Date of Event
September 1, 2014
Report Date
September 1, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED. NO MOISTURE DAMAGE WAS FOUND INSIDE OF THE INSULIN PUMP. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT A DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS AND A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD HIGH BLOOD GLUCOSE LEVELS OF 227 MG/DL. THE CUSTOMER REPORTED A BUTTON ERROR ALARM FROM THE INSULIN PUMP. THE CUSTOMER ALSO REPORTED THAT THE INSULIN PUMP WAS EXPOSED TO WATER. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WOULD NEED TO BE REPLACED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO A BACK-UP PLAN PER THE HEALTH CARE PROVIDER'S INSTRUCTION. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
612400 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523NAH

Patients

Seq Age Sex Outcome Treatment
1 8 YR