FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4133328 · Received October 1, 2014

Report

Report Number
2032227-2014-31354
Event Type
Malfunction
Date Received
October 1, 2014
Date of Event
September 1, 2014
Report Date
September 1, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO SLIGHT MOISTURE DAMAGE TO THE KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED. NO TRACES OF MOISTURE WERE NOTED AT THE ELECTRONIC OR MOTOR ASSEMBLIES PER VISUAL INSPECTION. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED BELT CLIP SLOT AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

CUSTOMER REPORTED A BUTTON ERROR ON HIS INSULIN PUMP. CUSTOMER'S BLOOD GLUCOSE IS 129 MG/DL. CUSTOMER WENT IN THE POOL WITH THE DEVICE. CUSTOMER WAS ADVISED THE DEVICE NEEDS TO BE REPLACED, AND TO DISCONTINUE USE AND REVERT TO A BACK UP PLAN. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
611086 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 36 YR