PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-31350
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Date of Event
- September 1, 2014
- Report Date
- September 1, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
INSULIN PUMP RECEIVED WITH NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM DURING TESTING. INSULIN PUMP RECEIVED WITH MOISTURE DAMAGE ON THE ELECTRONICS, MOTOR, VIBRATOR, KEYPAD AND BATTERY TUBE ASSEMBLY. INSULIN PUMP RECEIVED WITHOUT THE BELT CLIP. INSULIN PUMP RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON LCD WINDOW, MISSING END CAP STICKER, CRACKED BATTERY TUBE THREADS AND CRACKED BELT CLIP SLOT.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. (B)(4).
IT WAS REPORTED THAT CUSTOMER'S INSULIN PUMP HAD A BUTTON ERROR. CUSTOMER STATED HER BELT CLIP BROKE TWO WEEKS AGO SO THE DEVICE HAS BEEN AGAINST THE SKIN. THE DEVICE WAS EXPOSED TO SWEAT. CUSTOMER'S BLOOD GLUCOSE WAS 5.7 MMOL/L. CUSTOMER WAS ADVISED THE DEVICE NEEDS TO BE REPLACED, AND TO DISCONTINUE USE AND REVERT TO A BACK UP PLAN. NOTHING FURTHER REPORTED.