FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4133280
·
Received October 1, 2014
Report
- Report Number
- 2032227-2014-31485
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Date of Event
- September 1, 2014
- Report Date
- September 1, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
INSULIN PUMP RECEIVED WITH CORRODED KEYPAD TRACES. NO DISPLAY RAMPING ON ITS OWN ANOMALY NOTED. ALL BUTTONS WORKING PROPERLY. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON LCD DISPLAY WINDOW.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT THE NUMBERS ON THE SCREEN OF THE INSULIN PUMP KEEP RAMPING AND THE KEYPAD IS UNRESPONSIVE. CUSTOMER DOES NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ISSUE. BLOOD GLUCOSE VALUE IS 185 MG/DL. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 610832 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723RNAP | A000290378 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |