PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-31464
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Date of Event
- September 1, 2014
- Report Date
- September 1, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- PATIENT
Narratives
THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED DURING TESTING. THE INSULIN PUMP HAD A CRACKED CASE AT A DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS AND A CRACKED RESERVOIR TUBE LIP.
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER REPORTED THAT THE INSULIN PUMP RECEIVED A BUTTON ERROR ALARM. BLOOD GLUCOSE VALUE IS 53 MG/DL; TREATED WITH GRAPES. CUSTOMER STATED HE THINKS HIS BLOOD GLUCOSE IS LOW BECAUSE OF HIS ACTIVITY. CUSTOMER STATED THAT THE DEVICE WAS EXPOSED TO SWEAT. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 611007 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723NAS | A000257056 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |