FDA Adverse Event
Malfunction
Summary report: N
S2S/SUR-FIT NATURA 2 PC - 2 PC UROSTOMY POUCH W/ACCUSEAL TAP
MDR report key: 4133012
·
Received September 29, 2014
Report
- Report Number
- 9618003-2014-11191
- Event Type
- Malfunction
- Date Received
- September 29, 2014
- Report Date
- December 17, 2012
- Manufacturer
- CONVATEC DOMINICAN REPUBLIC, INC.
- Product Code
- EXH
- PMA / PMN Number
- K840166
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
CONVATEC IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO THE FDA 483 ISSUED 04/16/2014. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFO RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FIELD ON A SUPPLEMENTAL MDR. ANALYSIS IS TAKEN FROM A TRENDING REPORT OF EVALUATIONS OR ADVERSE EVENTS FINALIZED 05/2010. IT WAS DETERMINED THAT THE ROOT CAUSE WAS TOO MUCH SILICONE ON THE TAP. THE INVESTIGATION SHOWS THAT THE ISSUE REPORTED AS "URINE WON'T DRAIN FROM POUCH" HAS BEEN CONFIRMED AND ACTIONS HAVE TAKEN PLACE TO ADDRESS IT.
Description of Event or Problem · 1
REPORT RECEIVED INDICATED THAT WHEN OPENING THE TAP, URINE DID NOT DRAIN. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 606947 | S2S/SUR-FIT NATURA 2 PC - 2 PC UROSTOMY POUCH W/ACCUSEAL TAP | URINARY, ILEOSTOMY | EXH | CONVATEC DOMINICAN REPUBLIC, INC. | 401544 | 2F01796 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |