ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-28032
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Report Date
- September 17, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/21/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP BLACK BOX SHOWED ONE POWER EVEN OCCURRING ON (B)(6) 2014; THE PUMP BATTERY VOLTAGE RECORDED AT THE TIME WAS ABOVE THE THRESHOLD FOR A BATTERY ALARM. DURING TESTING, THE PUMP BATTERY COMPARTMENT WAS NOTED TO BE CRACKED FROM THE CASE SEAL TO THE BUMPER PAD. THE PUMP BATTERY CAP THREADS WERE NOTED TO BE STRIPPED AND THE BATTERY CAP WAS UNABLE TO SECURE AND MAINTAIN ELECTRICAL CONNECTION. A TEST BATTERY CAP WAS INSERTED INTO THE PUMP AND THE PUMP POWERED ON NORMALLY AND NO FURTHER POWER INTERRUPTIONS OCCURRED DURING TESTING. THE PUMP COVER WAS REMOVED AND NO INTERMITTENT CONDITIONS WERE OBSERVED INSIDE THE PUMP. UNRELATED TO THE COMPLAINT, THE DISPLAY SCREEN WAS OBSERVED TO BE DIM AND DISCOLORED.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS ALLEGING THAT THE PUMP HAD AN ISSUE WITH INTERMITTENT POWER. THE REPORTER CONFIRMED THAT THERE WAS NO MOISTURE OF CORROSION IN THE PUMP AND CONFIRMED THAT THE BATTERY CAP WAS ABLE TO BE SECURED TO THE PUMP APPROPRIATELY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 612367 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR |