HOMECHOICE PRO
Report
- Report Number
- 1416980-2014-34016
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Date of Event
- July 3, 2014
- Report Date
- September 8, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EZ
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS COMPLETED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED IN THE EVENT HISTORY LOG REVIEW. THE CAUSE OF THE IIPV EVENT WAS DETERMINED TO BE A USE ERROR FURTHER SPECIFIED AS THE TOTAL TIDAL ULTRA FILTRATION REMOVAL WAS SET TOO LOW. THE HOMECHOICE APD SYSTEMS TRAINER¿S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 13:51:36. DURING NIGHT DRAIN CYCLE NINE, THE PATIENT'S ULTRAFILTRATION READING WAS 2157ML, INDICATING THE HOME PATIENT (HP) DRAINED 1757ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 612908 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |