PROMUS ELEMENT ?
Report
- Report Number
- 2134265-2014-05879
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Date of Event
- August 12, 2014
- Report Date
- September 9, 2014
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- NIQ
- PMA / PMN Number
- SIMILAR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CN
- Reporter Occupation
- PHYSICIAN
Narratives
AGE AT TIME OF EVENT: 18 YEARS OR OLDER DEVICE IS A COMBINATION PRODUCT. DEVICE EVALUATED BY MFR.:A VISUAL EXAMINATION OF THE CRIMPED STENT FOUND STENT STRUTS ON THE PROXIMAL PORTION OF THE CRIMPED STENT WERE DISTORTED, DEFORMED AND LIFTED UPWARDS FROM THE STENT PROFILE. IT IS LIKELY THE CRIMPED STENT MAY HAVE ENCOUNTERED RESISTANCE & SUBSEQUENT DAMAGE DURING WITHDRAWAL ATTEMPTS. THE BUMPER TIP OF THE DEVICE SHOWED NO SIGNS OF DAMAGE. THE BALLOON CONES PROFILES WERE REVIEWED AND NO ISSUES WERE NOTED WITH THE OVERALL BALLOON PROFILE. THE BALLOON WINGS WERE TIGHTLY WRAPPED AND EVENLY FOLDED AND WERE NOT SUBJECTED TO POSITIVE PRESSURE. A VISUAL AND TACTILE EXAMINATION FOUND MULTIPLE KINKS ALONG THE HYPOTUBE SHAFT. THIS TYPE OF DAMAGE IS CONSISTENT WITH EXCESSIVE FORCE BEING APPLIED TO THE DELIVERY SYSTEM. A VISUAL AND TACTILE EXAMINATION FOUND NO ISSUES WITH THE PROFILE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).
REPORTABLE BASED ON ANALYSIS COMPLETED ON (B)(4) 2014. IT WAS REPORTED THAT CROSSING DIFFICULTIES WERE ENCOUNTERED. A 2.50X32MM PROMUS ELEMENT ¿ DRUG-ELUTING STENT WAS ADVANCED TO TREAT THE TARGET LESION BUT FAILED TO CROSS THE LESION. THE PROCEDURE WAS THEN COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED STENT DAMAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 612530 | PROMUS ELEMENT ? | STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC - GALWAY | H7493911332250 | 0016515953 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |