FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4132549 · Received October 1, 2014

Report

Report Number
2032227-2014-31418
Event Type
Malfunction
Date Received
October 1, 2014
Date of Event
September 1, 2014
Report Date
September 1, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.(B)(4)

Additional Manufacturer Narrative · 1

INSULIN PUMP HAD MOISTURE DAMAGE ON KEYPAD TRACES. ALL BUTTONS FUNCTION PROPERLY. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS AND CRACKED RESERVOIR TUBE LIP. MOISTURE DAMAGE WAS FOUND ON ELECTRONIC AND MOTOR ASSEMBLY.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP'S ACT AND ESCAPE BUTTONS WERE NO LONGER RESPONDING. CUSTOMER REPORTED THAT SHE HAD RECENTLY JUMPED INTO A LAKE WHILE WEARING THE INSULIN PUMP. BLOOD GLUCOSE LEVEL WAS 256 MG/DL AT THE TIME OF THE CALL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
610394 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723NAH

Patients

Seq Age Sex Outcome Treatment
1 44 YR