FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4132489 · Received October 1, 2014

Report

Report Number
2032227-2014-31705
Event Type
Malfunction
Date Received
October 1, 2014
Date of Event
September 1, 2014
Report Date
September 2, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

UNIT RECEIVED WITH FULL SEGMENT DISPLAY AND CONSTANT TONE DUE TO FAULTY CAPACITOR ON THE INTERFACE BOARD. NO BLANK DISPLAY NOTED. UNABLE TO PERFORM OPERATING CURRENTS DUE TO STUCK IN FULL SEGMENT DISPLAY. THE INSULIN PUMP RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS AND RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THE CUSTOMER RECEIVED A BLANK DISPLAY. CUSTOMER'S BLOOD GLUCOSE WAS 200 MG/DL. IT WAS ALSO REPORTED THERE WAS A HIGH PITCHED BEEP ON THE INSULIN PUMP. TROUBLESHOOTING FOUND THE BATTERY COMPARTMENT WAS NOT DAMAGED OR CORRODED. THE CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
611445 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAH

Patients

Seq Age Sex Outcome Treatment
1 17 YR