FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4132413 · Received October 1, 2014

Report

Report Number
2032227-2014-31247
Event Type
Malfunction
Date Received
October 1, 2014
Date of Event
September 1, 2014
Report Date
September 1, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED A MOTOR ERROR ON HER INSULIN PUMP WHEN TRYING TO DELIVERY A BOLUS. CUSTOMER'S BLOOD GLUCOSE WAS 205 MG/DL. THERE WERE NO SIGNIFICANT EVENTS LEADING TO THE ALARM. THE DEVICE WAS NOT EXPOSED TO A STRONG MAGNETIC FIELD. THE ALARM WAS CLEARED. CUSTOMER DOES NOT USE THE SENSOR FEATURE. CUSTOMER IS ABLE TO REWIND THE DEVICE. CUSTOMER WAS ADVISED THAT THE DEVICE NEEDS TO BE REPLACED, AND TO DISCONTINUE USE AND REVERT TO A BACK UP PLAN. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
612523 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723RNAS

Patients

Seq Age Sex Outcome Treatment
1 74 YR