PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-31237
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Date of Event
- September 1, 2014
- Report Date
- September 1, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- OTHER
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
THE INSULIN PUMP PASSED THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST, AND EXCESSIVE NO DELIVERY ALARM TESTS. NO MOTOR ERROR ALARM WAS NOTED AND THE MOTOR PASSED THE MOTOR TEST. THE INSULIN PUMP HAD A CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A CRACKED CASE AT THE DISPLAY WINDOW CORNERS.
CUSTOMER REPORTED THAT THE INSULIN PUMP RETURNED A MOTOR ERROR ALARM DURING RESERVOIR FILL. BLOOD GLUCOSE LEVEL WAS NOT PROVIDED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 613016 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-554WWL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |