TELIGEN
Report
- Report Number
- 2124215-2014-14468
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Date of Event
- July 20, 2014
- Report Date
- September 16, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Removal / Correction Number
- Z-0026-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY. EXTERNAL VISUAL INSPECTION OF THE DEVICE NOTED NO ANOMALIES. THE BATTERY VOLTAGE WAS LOWER THAN EXPECTED, BUT STILL SUPPORTED FULL DEVICE FUNCTION. A SERIES OF AUTOMATED ELECTRICAL/FUNCTIONAL TESTS WERE CONDUCTED AND NO ISSUES WITH THE DEVICE¿S ABILITY TO DELIVER THERAPY WERE OBSERVED; BASIC SENSING, PACING AND SHOCKING FUNCTIONS OF THE DEVICE WERE VERIFIED. REVIEW OF THE DEVICE MEMORY INDICATED THAT A VOLTAGE ALERT WAS RECORDED. USING HISTORICAL DAILY BATTERY VOLTAGE MEASUREMENT DATA, ENGINEERS DETERMINED THAT THIS DEVICE WAS DEMONSTRATING BEHAVIOR CONSISTENT WITH A HIGH CURRENT CONDITION ASSOCIATED WITH A COMPROMISED LOW VOLTAGE CAPACITOR CONNECTED TO THE DEVICE¿S BATTERY. LOW VOLTAGE CAPACITORS ARE USED IN THE DEVICE¿S HIGH VOLTAGE CHARGING OPERATION IN ORDER TO FACILITATE FAST CHARGE TIMES. MALFUNCTION OF THESE CAPACITORS RESULTED IN A HIGH CURRENT DRAIN, WHICH WAS DEPLETING THIS DEVICE¿S BATTERY FASTER THAN NORMAL. ON AUGUST 29, 2013, BOSTON SCIENTIFIC DISTRIBUTED A LETTER TO PHYSICIANS CONCERNING THIS ISSUE. THIS LETTER INFORMED PHYSICIANS THAT, IN A SUBSET OF DEVICES, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS DEVICE IS A PART OF THE IDENTIFIED POPULATION.
(B)(4). ADDITIONAL INFORMATION IS EXPECTING, THIS REPORT WILL BE UPDATED UPON RECEIPT OF ADDITIONAL DETAILS.
(B)(4).
(B)(4). THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
ADDITIONAL INFORMATION INDICATES THAT THIS PRODUCT WAS EXPLANTED. THE PATIENT WILL REMAIN IN THE HOSPITAL UNTIL A NEW DEVICE IS IMPLANTED. THE PATIENT'S CLINICAL CONDITION IS STABLE.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EMITTING TONES. THE DEVICE WAS INTERROGATED AND RECORDED A CODE 1003 INDICATIVE OF BATTERY VOLTAGE TOO LOW FOR THE PROJECTED REMAINING CAPACITY. THE PHYSICIAN ELECTED TO HOSPITALIZE THE PATIENT UNTIL THE REPLACEMENT PROCEDURE IS COMPLETED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 613124 | TELIGEN | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | F110 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | F110 |