FDA Adverse Event Malfunction Summary report: N

ENERGEN

MDR report key: 4132221 · Received October 1, 2014

Report

Report Number
2124215-2014-15915
Event Type
Malfunction
Date Received
October 1, 2014
Date of Event
May 10, 2014
Report Date
July 24, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH OUT OF RANGE PACING IMPEDANCE MEASUREMENTS. THERE WAS A CONCERN THE LV LEAD HAD FRACTURED. THE LEAD WAS REPROGRAMMED TO UNIPOLAR WITH APPROPRIATE SENSING AND CAPTURE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
612558 ENERGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWP GUIDANT CRM CLONMEL IRELAND N141

Patients

Seq Age Sex Outcome Treatment
1 59 YR H210| N141| 4537| 4136| 4542| 0137