FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4132160 · Received October 1, 2014

Report

Report Number
2032227-2014-30987
Event Type
Injury
Date Received
October 1, 2014
Date of Event
September 1, 2014
Report Date
September 1, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. (B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE REWIND, BASIC OCCLUSION, PRIME AND DISPLACEMENT TESTS. UNIT RECEIVED WITH STUCK MOTOR ERROR ALARM LOOP DURING BOLUS DELIVERY. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. UNIT HAD CRACKED DISPLAY WINDOW, CRACKED CASE NEAR DISPLAY WINDOW CORNER, RESERVOIR TUBE, RESERVOIR TUBE WINDOW, RESERVOIR TUBE LIP, BELT CLIP SLOT, BATTERY TUBE THREADS, STAINED ADDRESS AND SERIAL NUMBER LABEL AND MISSING END CAP STICKER.

Description of Event or Problem · 1

CUSTOMER REPORTED HAVING RECENT HIGH BLOOD GLUCOSE LEVELS. TROUBLESHOOTING SHOWED THAT THE INSULIN PUMP WAS FUNCTIONING PROPERLY. BLOOD GLUCOSE LEVEL WAS 400 MG/DL AT THE TIME OF THE CALL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
612481 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522WWP

Patients

Seq Age Sex Outcome Treatment
1