PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-30985
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Date of Event
- September 1, 2014
- Report Date
- September 1, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FI
- Reporter Occupation
- OTHER
Narratives
THE INSULIN PUMP ALARMED DURING THE SELF TEST DUE TO A FAULTY CONNECTOR ON THE RADIO FREQUENCY BOARD. THE INSULIN PUMP HAD CRACKED BATTERY TUBE THREADS AND A CRACKED RESERVOIR TUBE LIP.
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4). THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
CUSTOMER'S NURSE REPORTED THAT THE INSULIN PUMP RETURNED SEVERAL ERROR ALARMS. CUSTOMER DID NOT RECALL ANY SIGNIFICANT EVENTS LEADING UP TO THE ERROR ALARMS. BLOOD GLUCOSE LEVEL WAS NOT PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 613015 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-554WWH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |