FDA Adverse Event Injury Summary report: N

ACUITY

MDR report key: 4132146 · Received October 1, 2014

Report

Report Number
2124215-2014-17158
Event Type
Injury
Date Received
October 1, 2014
Date of Event
July 23, 2014
Report Date
July 23, 2014
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY THIS LEAD HAS BEEN RETURNED. BOSTON SCIENTIFIC HAS CONCLUDED IT IS UNLIKELY THAT LEAD CHARACTERISTICS CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD EXHIBITED LOSS OF CAPTURE (LOC) WITH ASYSTOLE OF LESS THAN 2 SECONDS. THE LEAD WAS FOUND TO BE DISLODGED UNDER A CHEST X-RAY. THE PATIENT UNDERWENT A SURGICAL INTERVENTION WHERE THIS LEAD WAS EXPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
611962 ACUITY IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4554

Patients

Seq Age Sex Outcome Treatment
1 71 YR Hospitalization| L| R 4469| 4554| V273| 4087