FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4132015 · Received October 1, 2014

Report

Report Number
2032227-2014-30885
Event Type
Malfunction
Date Received
October 1, 2014
Date of Event
August 31, 2014
Report Date
September 1, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S RESERVOIR WAS EMPTY BUT STILL SHOWED 4 UNIT OF INSULIN. CUSTOMER ADVISED THAT THERE WAS NO ALARM TO INDICATE THAT THE RESERVOIR WAS EMPTY. TROUBLESHOOTING OCCURED. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME 172MG/DL. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
611230 530G INSULIN PUMP OZO OZO MEDTRONIC MINIMED MMT-751NAL

Patients

Seq Age Sex Outcome Treatment
1 63 YR