PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-31440
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Date of Event
- August 30, 2014
- Report Date
- September 1, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT
Narratives
INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM NOTED. INSULIN PUMP RECEIVED WITH MOISTURE DAMAGE NOTED ON MOTOR. INSULIN PUMP RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON LCD WINDOW AND CRACKED BATTERY TUBE THREADS.
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER REPORTED A BUTTON ERROR ALARM. THE CUSTOMER'S BLOOD GLUCOSE WAS 220 MG/DL. THE CUSTOMER STATED HE HAD DROPPED HIS INSULIN PUMP INTO THE POOL PRIOR TO THE ALARM OCCURRING. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION IS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 611293 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR |