PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4131880 · Received October 1, 2014

Report

Report Number
2032227-2014-31268
Event Type
Malfunction
Date Received
October 1, 2014
Date of Event
August 30, 2014
Report Date
December 23, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH NO UP BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE INSULIN HAD MINOR SCRATCHED LCD WINDOW, CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP AND STAINED END CAP STICKER.

Description of Event or Problem · 1

THE IT WAS REPORTED THAT THE BUTTONS ON THE CUSTOMER'S INSULIN PUMP WERE NOT RESPONDING DURING A BOLUS DELIVERY ATTEMPT. HER BLOOD GLUCOSE LEVEL WAS 173 MG/DL. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE PUMP AND REVERT TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
611867 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523NAP

Patients

Seq Age Sex Outcome Treatment
1 63 YR