FDA Adverse Event Malfunction Summary report: N

ROTATING MULTIPLE CLIP APPLIER

MDR report key: 4131879 · Received October 1, 2014

Report

Report Number
3005075853-2014-06782
Event Type
Malfunction
Date Received
October 1, 2014
Date of Event
September 16, 2014
Report Date
September 17, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. AT THE TIME OF THIS SUBMISSION, THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: DEVICE WAS RETURNED FOR ANALYSIS WITH A CLIP IN THE JAWS; THE CLIP WAS REMOVED IN ORDER TO MEASURE JAWS WIDTH AND THEY WERE FOUND TO BE YIELDED. A BAG WITH FOUR SCISSORED CLIPS WAS RECEIVED ALONG WITH THE INSTRUMENT. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED AND FED AND FORMED FOURTEEN SCISSORED CLIPS. IN ADDITION THE DEVICE LOCKED OUT AS INTENDED. POSSIBLE CAUSES FOR THE FOUND CONDITION OF THE YIELDED JAWS MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. ALTHOUGH IT IS NOT POSSIBLE TO CONCLUDE HOW THE CIRCUMSTANCES OCCURRED, IT IS KNOWN FROM THE HISTORY OF THE INSTRUMENT THAT AN INCORRECT/EXCESSIVE APPLICATION OF TORQUE TO THE JAWS DURING INSTRUMENT USE CREATES A MISALIGNMENT OF THE TIPS. IN ADDITION AS PER THE INSTRUCTIONS FOR USE ¿DO NOT EXCESSIVELY TWIST OR TORQUE THE INSTRUMENT JAWS WHEN POSITIONING THE INSTRUMENT ON A VESSEL AND FIRING. EXCESSIVE TWISTING OR TORQUING MAY RESULT IN CLIP MALFORMATION¿. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, SCISSORING OCCURRED AT THE 2ND FIRING. THE DEVICE WAS USED ON THE BLOOD VESSEL. THERE MIGHT HAVE BEEN TORQUING OR TWISTING OF THE DEVICE PRESENT. ANOTHER COMPETITIVE DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
611364 ROTATING MULTIPLE CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1