FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4131852 · Received October 1, 2014

Report

Report Number
2032227-2014-31496
Event Type
Malfunction
Date Received
October 1, 2014
Date of Event
August 30, 2014
Report Date
September 1, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP HAD AN ARROW BUTTON THAT DID NOT RESPOND DUE TO CORRODED KEYPAD TRACES. NO SCROLLING NUMBERS DISPLAY ANOMALY NOTED. NO MOISTURE DAMAGE ON ELECTRONICS NOTED. NO DISCOLORATION ON SCREEN NOTED. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, CRACKED RESERVOIR TUBE AND BLEEDING LCD.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER SWEATS MORE THAN THE NORMAL PERSON. CUSTOMER NOTICED ABOUT THREE WEEKS AGO THAT THE SCREEN WAS DISCOLORED. CUSTOMER CANNOT SCROLL DOWN TO REWIND AND THE ARROW BUTTONS DO NOT WORK. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 427 MG/DL. CUSTOMER STATED THAT THE PUMP WAS NOT WORKING AT ALL. INSULIN PUMP WILL NEED TO BE REPLACED. NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
612043 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAB

Patients

Seq Age Sex Outcome Treatment
1 58 YR