FDA Adverse Event Injury Summary report: N

FLEXTEND

MDR report key: 4131771 · Received October 1, 2014

Report

Report Number
2124215-2014-15304
Event Type
Injury
Date Received
October 1, 2014
Date of Event
June 30, 2014
Report Date
July 9, 2014
Manufacturer
CPI - DEL CARIBE
Product Code
DTB
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. VISUAL INSPECTION OF THE LEAD NOTED SETSCREW MARKS WERE NOTED ON EACH TERMINAL PIN AND RING. THERE WAS ALSO BLOOD INFILTRATION NOTED IN THE HELIX HOUSING. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INNER INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT WITH THIS RIGHT VENTRICULAR (RV) LEAD EXPERIENCED SYNCOPE AND RETURNED TO THE HOSPITAL TEN DAYS POST-IMPLANT. ELECTROGRAMS REVEALED LOSS OF CAPTURE ON THE RV LEAD. INVESTIGATION PERFORMED BY THE CATHETER LAB DID NOT FIND ANY EVIDENCE THAT THE LEAD HAD DISLODGED. A LEAD REVISION WAS PERFORMED AND THE RV LEAD WAS TESTED ON THE PACING SYSTEM ANALYZER (PSA). BOTH THE PACING THRESHOLD AND IMPEDANCE MEASUREMENTS WERE FOUND TO BE INITIALLY HIGH WHEN THE LEAD WAS REPOSITIONED; HOWEVER, THE PACING IMPEDANCE MEASUREMENTS WERE NOTED TO BE OUT OF RANGE. THE RV LEAD WAS REPOSITIONED SEVERAL TIMES WITH THE SAME RESULT. THIS RV LEAD WAS EXPLANTED AND SUCCESSFULLY REPLACED WITH A NEW RV LEAD OF THE SAME MODEL. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. THE LEAD WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
612941 FLEXTEND IMPLANTABLE LEAD DTB CPI - DEL CARIBE 4088

Patients

Seq Age Sex Outcome Treatment
1 61 YR Hospitalization| L| R 4087| S208| 4088