FDA Adverse Event Malfunction Summary report: N

INSIGNIA

MDR report key: 4131722 · Received October 1, 2014

Report

Report Number
2124215-2014-15087
Event Type
Malfunction
Date Received
October 1, 2014
Date of Event
July 8, 2014
Report Date
September 9, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
PMA / PMN Number
D970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING THE EXPLANT OF THIS PATIENT'S PACEMAKER FOR NORMAL BATTERY DEPLETION (NBD) THE PATIENTS LEAD WAS FOUND TO BE STUCK IN THE DEVICE HEADER. THE LEAD WAS DAMAGED IN AN ATTEMPT TO FREE IT FROM THE DEVICE, AND WAS UNABLE TO BE RE-USED. THE LEAD WAS SURGICALLY ABANDONED AND BOTH THE LEAD AND DEVICE WERE SUCCESSFULLY REPLACED. TO DATE, NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
613187 INSIGNIA IMPLANTABLE PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND 1298

Patients

Seq Age Sex Outcome Treatment
1 1204 MO 1298| K063| 1283| 294-05| 430-07| 4136| MISMATCH