INSIGNIA
Report
- Report Number
- 2124215-2014-15087
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Date of Event
- July 8, 2014
- Report Date
- September 9, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- PMA / PMN Number
- D970003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING THE EXPLANT OF THIS PATIENT'S PACEMAKER FOR NORMAL BATTERY DEPLETION (NBD) THE PATIENTS LEAD WAS FOUND TO BE STUCK IN THE DEVICE HEADER. THE LEAD WAS DAMAGED IN AN ATTEMPT TO FREE IT FROM THE DEVICE, AND WAS UNABLE TO BE RE-USED. THE LEAD WAS SURGICALLY ABANDONED AND BOTH THE LEAD AND DEVICE WERE SUCCESSFULLY REPLACED. TO DATE, NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 613187 | INSIGNIA | IMPLANTABLE PULSE GENERATOR | LWP | GUIDANT CRM CLONMEL IRELAND | 1298 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 1204 MO | 1298| K063| 1283| 294-05| 430-07| 4136| MISMATCH |