FDA Adverse Event Malfunction Summary report: N

INSIGNIA

MDR report key: 4131677 · Received October 1, 2014

Report

Report Number
2124215-2014-17557
Event Type
Malfunction
Date Received
October 1, 2014
Date of Event
July 15, 2014
Report Date
July 15, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
PMA / PMN Number
D970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING THE ELECTIVE PROCEDURE TO REPLACE THIS PATIENT'S DEVICE, THE ASSOCIATED RIGHT VENTRICULAR (RV) LEAD WAS STUCK IN THE DEVICE HEADER. THE LEAD WAS NOTED TO HAVE SIGNIFICANT INSULATION COMPROMISE IN THE POCKET AND THE TERMINAL PIN AND RING ASSEMBLY HAD COME UNDONE. THERE WERE NO NOTED ABNORMALITIES IN LEAD PERFORMANCE PRIOR TO CHANGEOUT. THE DEVICE AND LEAD WERE REMOVED FROM SERVICE AND REPLACED WITHOUT FURTHER INCIDENT. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
612141 INSIGNIA IMPLANTABLE PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND 1298

Patients

Seq Age Sex Outcome Treatment
1 83 YR 1283| 438-01| K064| 4076| 294-03| 1298