FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 4131393 · Received October 1, 2014

Report

Report Number
1525712-2014-06525
Event Type
Malfunction
Date Received
October 1, 2014
Report Date
September 12, 2014
Manufacturer
UNKNOWN
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

PER DEALER SIEVE BEDS ARE CAUSING THE UNIT TO ALARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
610617 PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) 868.5440 CAW UNKNOWN CONCENTRATOR

Patients

Seq Age Sex Outcome Treatment
1 Other