CONTOUR CUTTER STAPLER
Report
- Report Number
- 3005075853-2014-06772
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Report Date
- September 17, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K040038
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. AT THE TIME OF THIS SUBMISSION, THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS. ADDITIONAL INFORMATION WAS REQUESTED AND THE FOLLOWING WAS OBTAINED: DID THE DEVICE CUT? YES. IF YES, WAS THE CUT LINE COMPLETE? YES. WERE THERE STAPLES DEPLOYED, BUT THEY WERE MISSING ON THE RECTAL SIDE? NO STAPLES.
(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS RESULTS SHOWED THAT THE CS40G DEVICE WAS RECEIVED IN GOOD VISUAL CONDITIONS AND WITH A CARTRIDGE RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED VOID OF STAPLES, WITH THE WASHER COMPLETELY CUT, WITH THE DRIVERS EXPOSED AND WITH THE KNIFE RECESS BELOW THE CARTRIDGE DECK. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH AN ENGINEERING SAMPLE RELOAD AND THE DEVICE FIRED, FORMING ALL STAPLES AND CUTTING AS INTENDED. THE CUT LINE WAS COMPLETE, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE DEVICE LOCKOUT AND THE RELOAD LOCKOUT WERE FUNCTIONAL. A BATCH RECORD REVIEW WAS PERFORMED AND THE BATCH HAD NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A RECTUM PROCEDURE, THERE WERE NO CLIPS ON THE RECTAL SIDE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 611993 | CONTOUR CUTTER STAPLER | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |