FDA Adverse Event Malfunction Summary report: N

STRAIGHT REDUCTION CLAMP

MDR report key: 4131295 · Received October 1, 2014

Report

Report Number
0008031020-2014-00456
Event Type
Malfunction
Date Received
October 1, 2014
Date of Event
September 9, 2014
Report Date
September 11, 2014
Manufacturer
STRYKER TRAUMA SELZACH
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE REPORTED EVENT OF THE POINTS DEFORM AND BREAK OFF FOR STRAIGHT REDUCTION CLAMP COULD BE CONFIRMED. BASED ON THE INVESTIGATION, THE ROOT CAUSE ZAEHNE SPITZ BENDING WAS ATTRIBUTED TO A USER RELATED ISSUE. IN REGARD OF THE RESULTS ON BML 12-169: VARIAX CLAVICLE REDUCTION CLAMP: CORROSION AND CYCLIC STRENGTH TEST [2] AND MEMO V12073_500: VARIAX CLAVICLE CLAMP AND FORCEPS: CYCLIC STRENGTH TEST [3]. WE CAN CONCLUDE THAT THE ROOT CAUSE OF THE FAILURE MODE REF#703822 STRAIGHT REDUCTION CLAMP - ZAEHNE SPITZ BENDING (TIP BENDING) IS DUE TO ¿USER RELATED¿. THE STRAIGHT REDUCTION CLAMP SHALL BE USE ONLY ON BONE MATERIAL. IN THE EVENT OF A USE IN A MATERIAL WITH A YIELD STRENGTH SUPERIOR AS THE BONE THE FAILURE MODE ZAEHNE SPITZ BENDING COULD OCCUR. [2] MEMO V12073_500: VARIAX CLAVICLE CLAMP AND FORCEPS: CYCLIC STRENGTH TEST [3] BML 12-169: VARIAX CLAVICLE REDUCTION CLAMP: CORROSION AND CYCLIC STRENGTH TEST A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. NO INDICATIONS OF MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE FOUND DURING THE INVESTIGATION.

Description of Event or Problem · 1

IT IS REPORTED BY THE NURSE, THAT THE LOBSTER CLAMP DOES NOT CLOSE UNDER TENSION OF A REDUCTION. THE WEBER CLAMP POINTS DEFORM AND BREAK OFF DURING BONE REDUCTION.

Description of Event or Problem · 1

IT IS REPORTED BY THE NURSE, THAT THE LOBSTER CLAMP DOES NOT CLOSE UNDER TENSION OF A REDUCTION. THE WEBER CLAMP POINTS DEFORM AND BREAK OFF DURING BONE REDUCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
611876 STRAIGHT REDUCTION CLAMP ORTHOPEDIC MANUAL SURGICAL INSTRUMENT LXH STRYKER TRAUMA SELZACH R37871

Patients

Seq Age Sex Outcome Treatment
1 Other