STRAIGHT REDUCTION CLAMP
Report
- Report Number
- 0008031020-2014-00456
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Date of Event
- September 9, 2014
- Report Date
- September 11, 2014
- Manufacturer
- STRYKER TRAUMA SELZACH
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- NURSE
Narratives
ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.
EVALUATION SUMMARY: THE REPORTED EVENT OF THE POINTS DEFORM AND BREAK OFF FOR STRAIGHT REDUCTION CLAMP COULD BE CONFIRMED. BASED ON THE INVESTIGATION, THE ROOT CAUSE ZAEHNE SPITZ BENDING WAS ATTRIBUTED TO A USER RELATED ISSUE. IN REGARD OF THE RESULTS ON BML 12-169: VARIAX CLAVICLE REDUCTION CLAMP: CORROSION AND CYCLIC STRENGTH TEST [2] AND MEMO V12073_500: VARIAX CLAVICLE CLAMP AND FORCEPS: CYCLIC STRENGTH TEST [3]. WE CAN CONCLUDE THAT THE ROOT CAUSE OF THE FAILURE MODE REF#703822 STRAIGHT REDUCTION CLAMP - ZAEHNE SPITZ BENDING (TIP BENDING) IS DUE TO ¿USER RELATED¿. THE STRAIGHT REDUCTION CLAMP SHALL BE USE ONLY ON BONE MATERIAL. IN THE EVENT OF A USE IN A MATERIAL WITH A YIELD STRENGTH SUPERIOR AS THE BONE THE FAILURE MODE ZAEHNE SPITZ BENDING COULD OCCUR. [2] MEMO V12073_500: VARIAX CLAVICLE CLAMP AND FORCEPS: CYCLIC STRENGTH TEST [3] BML 12-169: VARIAX CLAVICLE REDUCTION CLAMP: CORROSION AND CYCLIC STRENGTH TEST A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. NO INDICATIONS OF MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE FOUND DURING THE INVESTIGATION.
IT IS REPORTED BY THE NURSE, THAT THE LOBSTER CLAMP DOES NOT CLOSE UNDER TENSION OF A REDUCTION. THE WEBER CLAMP POINTS DEFORM AND BREAK OFF DURING BONE REDUCTION.
IT IS REPORTED BY THE NURSE, THAT THE LOBSTER CLAMP DOES NOT CLOSE UNDER TENSION OF A REDUCTION. THE WEBER CLAMP POINTS DEFORM AND BREAK OFF DURING BONE REDUCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 611876 | STRAIGHT REDUCTION CLAMP | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | LXH | STRYKER TRAUMA SELZACH | R37871 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |