FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 4131294 · Received October 1, 2014

Report

Report Number
1644487-2014-02523
Event Type
Malfunction
Date Received
October 1, 2014
Date of Event
July 31, 2014
Report Date
September 8, 2014
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

IT WAS INITIALLY REPORTED THAT THE PHYSICIAN WAS HAVING ISSUES GETTING HIS HANDHELD DEVICE TO RESPOND ON (B)(6) 2014. A HARD RESET WAS PERFORMED WHICH RESOLVED THE ISSUES WITH THE HANDHELD THE SAME DAY. THE PHYSICIAN AGAIN EXPERIENCED DIFFICULTIES GETTING HIS HANDHELD TO RESPOND ON (B)(6) 2014. THE HANDHELD WAS ONLY RESPONDING TO BUTTON PRESSES AND NOT TO SCREEN PRESSES. THE HANDHELD SCREEN WAS CLEANED AND HARD RESETS WERE PERFORMED; HOWEVER, THE ISSUE DID NOT RESOLVE. THE LOCK BUTTON WAS CONFIRMED TO BE DISENGAGED. FOLLOW-UP REVEALED THAT THE HANDHELD WOULD NOT PAST THE ALIGNMENT SCREEN. THE HANDHELD DEVICE AND SOFTWARE FLASHCARD HAVE BEEN RETURNED TO THE MANUFACTURER WHERE ANALYSIS IS CURRENTLY UNDERWAY.

Description of Event or Problem · 1

ANALYSIS OF THE RETURNED HANDHELD DEVICE AND SOFTWARE FLASHCARD WAS COMPLETED. IT WAS VERIFIED THAT THE HANDHELD WAS UNABLE TO ADVANCE PAST THE SCREEN ALIGNMENT UTILITY. THE CAUSE FOR THE DISPLAY ANOMALY IS ASSOCIATED WITH RESISTANCE VALUES BEING HIGHER THAN EXPECTED IN THE TOUCH SCREEN CIRCUITRY. NO FURTHER ANOMALIES ASSOCIATED WITH THE HANDHELD PERFORMANCE WERE IDENTIFIED DURING, THE ANALYSIS. NO ANOMALIES ASSOCIATED WITH FLASHCARD SOFTWARE OR DATABASES WERE IDENTIFIED DURING THE FLASHCARD ANALYSIS. THE FLASHCARD AND SOFTWARE PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
611413 PROGRAMMING SOFTWARE PROGRAMMING COMPUTER LYJ CYBERONICS, INC. MODEL 250 1073773

Patients

Seq Age Sex Outcome Treatment
1