FDA Adverse Event Malfunction Summary report: N

QUANTA SYSTEM S.P.A.

MDR report key: 4131229 · Received September 29, 2014

Report

Report Number
3004378299-2014-00081
Event Type
Malfunction
Date Received
September 29, 2014
Date of Event
September 3, 2014
Report Date
September 23, 2014
Manufacturer
QUANTA SYSTEM, S.P.A.
Product Code
GEX
PMA / PMN Number
K091909
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PROBLEM WAS DUE TO AN INTERNAL COMPONENT FAILURE. AFTER THE REPLACEMENT OF THIS COMPONENT, THE LASER SYSTEM RESTARTED TO WORK. WE ARE UNAWARE ABOUT ANY PATIENT INJURY.

Description of Event or Problem · 1

THE LASER SYSTEM HAS THE FOLLOWING PROBLEM: LOW OUTPUT POWER DURING INSTALLATION. NO ADVERSE EFFECTS TO THE PATIENT WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
606400 QUANTA SYSTEM S.P.A. SURGICAL LASER GEX QUANTA SYSTEM, S.P.A. LITHO

Patients

Seq Age Sex Outcome Treatment
1