FDA Adverse Event Malfunction Summary report: N

QUANTA SYSTEM S.P.A.

MDR report key: 4131201 · Received September 29, 2014

Report

Report Number
3004378299-2014-00080
Event Type
Malfunction
Date Received
September 29, 2014
Date of Event
September 15, 2014
Report Date
September 22, 2014
Manufacturer
QUANTA SYSTEM, S.P.A.
Product Code
GEX
PMA / PMN Number
K131081
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PROBLEM WAS DUE TO AN INTERNAL COMPONENT FAILURE. AFTER THE REPLACEMENT OF THIS COMPONENT, THE LASER SYSTEM RESTARTED TO WORK. WE ARE UNAWARE ABOUT PATIENT INJURY, BROKEN COMPONENT FOUND DURING INSTALLATION.

Description of Event or Problem · 1

THE LASER SYSTEM HAS THE FOLLOWING PROBLEM: OUTPUT POWER LOWER THAN EXPECTED DURING INSTALLATION. NO ADVERSE EFFECTS TO THE PATIENT WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
606218 QUANTA SYSTEM S.P.A. SURGICAL LASER GEX QUANTA SYSTEM, S.P.A. CYBER TM 200

Patients

Seq Age Sex Outcome Treatment
1