FDA Adverse Event
Malfunction
Summary report: N
QUANTA SYSTEM S.P.A.
MDR report key: 4131201
·
Received September 29, 2014
Report
- Report Number
- 3004378299-2014-00080
- Event Type
- Malfunction
- Date Received
- September 29, 2014
- Date of Event
- September 15, 2014
- Report Date
- September 22, 2014
- Manufacturer
- QUANTA SYSTEM, S.P.A.
- Product Code
- GEX
- PMA / PMN Number
- K131081
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PROBLEM WAS DUE TO AN INTERNAL COMPONENT FAILURE. AFTER THE REPLACEMENT OF THIS COMPONENT, THE LASER SYSTEM RESTARTED TO WORK. WE ARE UNAWARE ABOUT PATIENT INJURY, BROKEN COMPONENT FOUND DURING INSTALLATION.
Description of Event or Problem · 1
THE LASER SYSTEM HAS THE FOLLOWING PROBLEM: OUTPUT POWER LOWER THAN EXPECTED DURING INSTALLATION. NO ADVERSE EFFECTS TO THE PATIENT WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 606218 | QUANTA SYSTEM S.P.A. | SURGICAL LASER | GEX | QUANTA SYSTEM, S.P.A. | CYBER TM 200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |