ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2014-00788
- Event Type
- Malfunction
- Date Received
- September 30, 2014
- Date of Event
- September 2, 2014
- Report Date
- September 3, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RT345 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: THE COMPLAINT RT345 ADULT BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) AND WAS VISUALLY INSPECTED, PRESSURE TESTED AND IMMERSED IN A WATER BATH TO FIND THE SOURCE OF THE LEAK. RESULTS: VISUAL INSPECTION REVEALED THAT THE EVAQUA (EXPIRATORY) LIMB PROXIMAL CONNECTOR WAS INCORRECTLY ASSEMBLED. THE PRESSURE TEST SHOWED THAT THE CIRCUIT WAS OUT OF SPECIFICATION. THE WATER BATH TEST CONFIRMED THAT THE PROXIMAL CONNECTOR ON THE EXPIRATORY LIMB WAS THE SOURCE OF THE LEAK. CONCLUSION: ALL BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED BEFORE RELEASING FOR DISTRIBUTION, AND THOSE THAT FAIL ARE REJECTED. THE USER INSTRUCTIONS SUPPLIED WITH ADULT EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT. SET APPROPRIATE VENTILATOR ALARMS. HOSPITAL STAFF CORRECTLY CHECKED THE BREATHING CIRCUIT BEFORE PATIENT USE WHICH IS IN LINE WITH OUR USER INSTRUCTIONS.
A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT AN RT345 ADULT BREATHING CIRCUIT FAILED THE SERVO VENTILATOR LEAK TEST BEFORE USE ON A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 608894 | ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT345 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | SERVO VENTILATOR |