ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-27970
- Event Type
- Injury
- Date Received
- September 30, 2014
- Report Date
- September 27, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 01/20/2015 WITH THE FOLLOWING FINDINGS: BLACK BOX SHOWS UNACKNOWLEDGED ¿CS088¿ ALARMS FOR 4HR ON (B)(6) 2014, RETURNED BATTERY CAP AND TEST CARTRIDGE CAP WERE USED TO COMPLETE THE INVESTIGATION. TOTAL DAILY DOSE ADDS UP AND EQUALS THE PROGRAMMED BASAL RATE TARGET. NO ¿CS088¿ ALARM OR ANY EAW OCCURRED DURING THE INVESTIGATION, THE PRODUCT PERFORMS WITHIN SPECIFICATIONS. THE PUMP REFLECTED 24U AFTER A 24HR ON 1U/HR DURATION TEST, UNABLE TO DUPLICATE ¿CS088¿ COMPLAINT, THE PUMP¿S COVER WAS REMOVED, NO INTERMITTENT CONDITION WAS FOUND TO THE POWER PCB. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE REPORTER CONTACTED ANIMAS ON (B)(6) 2014 ALLEGING A CALL SERVICE 088 SERVICE ALARM ISSUE. THE REPORTER STATED THAT THE PATIENT HAD A BLOOD GLUCOSE (BG) OF 29.5MMOL/L WITH POLYURIA, POLYDIPSIA AND NAUSEA. THE PATIENT DID NOT RECEIVE ANY TREATMENT ABOVE AND BEYOND THE USUAL ROUTINE CARE OF DIABETES MANAGEMENT. THIS REPORT IS BEING MADE DUE TO THE BG EXCURSION THE PATIENT EXPERIENCED DUE TO AN ALLEGED CALL SERVICE ALARM ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 609366 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR | Life Threatening |