ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-27947
- Event Type
- Malfunction
- Date Received
- September 30, 2014
- Report Date
- September 19, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 10/13/2015. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/24/2015 WITH THE FOLLOWING FINDINGS: BLACK BOX SHOWS DATES FROM 8/16/2015 -8/25/2015. BLACK BOX DATA FROM DATE OF COMPLAINT HAS BEEN OVERWRITTEN HOWEVER CURRENT BB SHOWS ONE EAW 162 LOSS OF PRIME WARNING WITH A NON ZERO LOW FORCE ON 8/23/2015. THE PUMP POWERS WITH RETURNED BATTERY CAP. EXERCISED PUMP FOR 24 HOURS WITH A 1 UNIT PER HOUR BASAL PROGRAM. NO LOSS OF PRIME OR PRIME RELATED WARNINGS DUPLICATED. FORCE CALIBRATION CHECK PASSES WITH A READING OF 170. REMOVED PUMP CASE. FORCE SENSOR PINS SEATED AND SOLDER CONNECTIONS INTACT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 607986 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR |