FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 4130858 · Received September 30, 2014

Report

Report Number
2938836-2014-16267
Event Type
Injury
Date Received
September 30, 2014
Date of Event
August 14, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION: ANALYSIS WAS NORMAL. NO ANOMALY WAS FOUND. A LEAD TIP STIFFNESS TEST WAS PERFORMED AND THE LEAD WAS FOUND TO BE WITHIN SPECIFICATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A FOLLOW-UP, AN INCREASE IN CAPTURE THRESHOLD WAS OBSERVED. AN X-RAY SHOWED PERFORATION. THE LEAD WAS EXPLANTED AND REPLACED. THE PATIENT WAS DOING WELL AFTER THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
607757 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7122/65 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention