FDA Adverse Event Malfunction Summary report: N

QUADRA ASSURA CRT-D, DF-4 CONNECTOR

MDR report key: 4130793 · Received September 30, 2014

Report

Report Number
2938836-2014-16254
Event Type
Malfunction
Date Received
September 30, 2014
Date of Event
September 12, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
P910023
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(6).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING FOLLOW-UP, POST-PACED T WAVE OVERSENSING ON THE VENTRICULAR CHANNEL WAS OBSERVED. PATIENT WAS ASYMPTOMATIC. PROGRAMMING CHANGES WERE MADE AND RESOLVED THE OVERSENSING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
607399 QUADRA ASSURA CRT-D, DF-4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC., CRMD CD3365-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 54 YR