FDA Adverse Event Malfunction Summary report: N

ELLIPSE DR, DF-4 CONNECTOR

MDR report key: 4130719 · Received September 30, 2014

Report

Report Number
2938836-2014-16226
Event Type
Malfunction
Date Received
September 30, 2014
Date of Event
September 10, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
LWS
PMA / PMN Number
P910023
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT NON-SUSTAINED LEAD NOISE DUE TO POST-PACED T WAVE OVERSENSING WAS OBSERVED VIA REMOTE TRANSMISSION. PATIENT WAS ASYMPTOMATIC. PROGRAMMING CHANGES WERE MADE AND RESOLVED THE OVERSENSING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
607299 ELLIPSE DR, DF-4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST. JUDE MEDICAL, INC., CRMD CD2311-36Q NA

Patients

Seq Age Sex Outcome Treatment
1 67 YR 7120Q/58, BNP031277