FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4130684 · Received September 30, 2014

Report

Report Number
2032227-2014-30772
Event Type
Malfunction
Date Received
September 30, 2014
Date of Event
August 31, 2014
Report Date
August 31, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

ALL BUTTONS FUNCTION PROPERLY. HOWEVER, INSULIN PUMP HAD MOISTURE DAMAGE WAS NOTED ON KEYPAD TRACES. INSULIN PUMP HAD MOISTURE DAMAGE NOTED ON ELECTRONIC ASSEMBLY. CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON DISPLAY WINDOW AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP HAD UNRESPONSIVE KEYPAD AND HAD MOISTURE/ FLUID DAMAGED. CUSTOMER'S BLOOD GLUCOSE WAS 144 MG/DL. CUSTOMER STATED HE JUMPED INTO THE POOL WITH THE INSULIN PUMP. CUSTOMER STATED THAT THERE WAS FLUID UNDER THE LIQUID CRYSTAL DISPLAY AND CRACK ON THE BODY OF THE INSULIN PUMP. . ADVISED CUSTOMER THE INSULIN PUMP WOULD BE REPLACED. ADVISED TO DISCONTINUE USING THE PUMP AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROVIDER. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
607268 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAB

Patients

Seq Age Sex Outcome Treatment
1 15 YR