FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4130678 · Received September 30, 2014

Report

Report Number
2032227-2014-30762
Event Type
Malfunction
Date Received
September 30, 2014
Date of Event
August 30, 2014
Report Date
August 31, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

NO BUTTON ERROR ALARM WAS NOTED. THE ACT AND ESCAPE BUTTONS DID NOT RESPOND DUE TO CORRODED KEYPAD TRACES. NO MOISTURE DAMAGE WAS NOTED ON THE ELECTRONICS. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, CRACKED CASE AT THE DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP AND MISSING END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER STATED WAS SWEATING A LOT PRIOR TO GETTING BUTTON ERROR. CUSTOMER'S BLOOD GLUCOSE WAS 141 MG/DL. ADVISED CUSTOMER THE INSULIN PUMP WOULD BE REPLACED. ADVISED TO DISCONTINUE USING THE PUMP AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROVIDER. NO FURTHER INFORMATION PROVIDED.