PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-28885
- Event Type
- Malfunction
- Date Received
- September 30, 2014
- Date of Event
- August 28, 2014
- Report Date
- August 28, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM DURING TESTING. INSULIN PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, CRACKED CASE AT THE DISPLAY WINDOW CORNERS, MINOR SCRATCHED LCD WINDOW, AND SCRATCHED RESERVOIR TUBE WINDOW.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER'S BLOOD GLUCOSE READING WAS 160 MG/DL. NO SIGNIFICANT EVENTS LEADING TO THE ALARM WERE OBSERVED. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.